AIP-600C Infuson Pump
- rate mode,drip mode,time mode,body weight mode
- Support a wide variety brands of infusion set
- Anti-bolus system to reduce significantly bolus after occlusion sudden release
- multi Visual and audible alarms system
- Product description: drug library optional, Infusion accuracy ±5% , Flow rate range 0.1-1200ml/h, Bolus rate 300-1200ml/h, KVO rate 1-5ml/h, 1500 history records storage
Product introduction
1. Arc design, compact & light weight
2. Peristaltic system, more accurate infusion
3. Optional drug library and intermittent infusion mode
4. 3 levels of occlusion detection, 4 levels of air bubble detection, more safer infusion
5. Purge & programmable Bolus function for time saving
6.Anti-bolus function, diminish the unwanted bolus volume after removal of the occlusion
7. Versatile alarm functions to alert the device operation status
8. Optional drop sensor
9. RS232 interface
Compact design
The compact and light weight design saves space and is beneficial during patient transfer
Large &colorful LCD display
Free flow protection
The built-in flow stop feature helps reduce the risk of accidental free flow prevent free flow
when pump door is open.
Anti-bolus function
Diminishes the volume of unwanted Bolus injected to the patient after removal of the
occlusion cause.Anti-reverse rotation detection
Panel lock
The panel lock feature helps prevent unauthorized changes of any instrument setting
Versatile functions
2000 history records,Real time display,RS232 interface (optional)
90°rotation pole clamp convenient for horizontal bar, vertical IV pole & platform use
User-friendly operation
Soft key design, easy to operate,Load directly the last infusion rate and volume limit
Adjustable buzzer volume (3 levels),Top indicator light reflects status as well as alarm case
Night mode (key tone off screen dimmed, top indicator light off) to ensure good rest of
patients removable pump body for easy cleaning
Rs232 interface
2000 event logs and alarm information can be transferred to computer for review
90°rotatable pole clamp Fixation on vertical pole or horizontal bar
Removable pump body Convenient cleaning
Mode |
AIP-600C infuson pump |
0.1~1200ml/h |
|
|
Micro mode: 0.1-99.9 ml/h in 0.1 ml increments |
Infusion mode |
Rate mode, drip mode, time mode, body weight mode |
Volume to be infused (VTBI) |
0~9999ml |
Volume infused (∑) |
0~36000ml |
Infusion accuracy |
±5% |
KVO rate |
1~5ml/h |
Bolus rate |
300~1200ml/h |
Purge |
600ml/h |
Air bubble detection |
ultrasound sensor detection with four levels of air in line alarm adjustable |
Occlusion pressure |
3 adjustable occlusion pressure settings: low, middle and high |
Alarms |
visual and audible alarms: Door open, Air-In-Line, |
Occlusion, Infusion completion, infusion near over, |
|
start reminder function, low battery, battery depleted, |
|
malfunction etc |
|
Computer interface |
Rs232 (optional) |
History records |
2000 history records |
Power supply |
AC:100~240V,50/60Hz DC:12V±1V |
Rechargeable lithium polymer battery, 7.4V, 1900mAh |
|
|
Can operate for more than 3 hours after fully charged at 25ml/h |
Battery recharge time |
10h with power on, 3h with power off |
Power consumption |
25VA |
Classification |
class I, type CF |
Waterproof level |
IP * 3 |
Mode of operation |
continuous |
Dimensions |
145 * 100 *120 mm (length * width * height) |
Weight |
≤1.4kg |
Temperature:5°~40°,relative humidity:10~95%, |
|
Operating conditions |
Atmosphere pressure: 86~106kPa |
Temperature:-20°~ +60°,relative humidity:10~95%, |
|
Transport conditions |
Atmosphere pressure: 50~106kPa |
Temperature:-20°~ +45°,relativehumidity:10~95%, |
|
Storage conditions |
Temperature:-20°~ +45°,relativehumidity:10~95%, |
Atmosphere pressure: 50~106kPa |